Reactor Technology

Technical Specifications: BR-1000/1500/2000
BR-1000 | BR-1500 | BR-2000 | ||
---|---|---|---|---|
General information | Material | Stainless Steel 316Ti (1.4571) or Hastelloy C-22 (2.4602) | ||
Temperature max. | 300° C | 300° C | 300° C | |
Pressure max. | 200 bar | 200 bar | 200 bar | |
Reactor Vessel | Volume | approx. 1,220 ml | approx. 1,740 ml | approx. 2,230 ml |
Inner diameter | 90 mm | 90 mm | 90 mm | |
Inner height | 193 mm | 275 mm | 352 mm | |
Weight | approx. 10.5 kg | approx. 13 kg | approx. 16 kg | |
TFM™-PTFE insert* (integrated) | Volume | approx. 1,000 ml | approx. 1,460ml | approx. 1,880 ml |
Inner diameter | 84 mm | 84 mm | 84 mm | |
Inner height | 182.5 mm | 264 mm | 340 mm | |
Armatures | Standard Armatures | Rupture Disc, Dip Tube, Valve, Tool | ||
Pressure Measurement | analog and/or digital | |||
Ports (total) | 6 | 6 | 6 | |
Ports (free) | 2** | 2** | 2** | |
Type of connection | 8 mm Tube Connection | |||
Heating Systems (optional) | Electrical (Hot Plate) | - | - | - |
Electrical (Heating Jacket) | ✓ | ✓ | ✓ | |
Via Fluid (Double Jacket) | ✓ | ✓ | ✓ | |
Via Fluid (Heating Coil) | ✓ | ✓ | ✓ | |
Stirring | Via Hot Plate | - | - | - |
RV-GL | ✓ | ✓ | ✓ |
* When using PTFE inserts the maximum allowable temperature is 230° C
** The number of free ports can be enlarged by the use of T-connections
Compare products
Range of Applications
Technology
- Syntheses like hydrogenation of C-C triple bonds, homogeneous/heterogeneous catalysis and polymerisation
- Catalyst researches like tests of catalysts consisting of liquid or solid materials
Automotive
- Material testing like aging tests of e.g. elastomers, ceramics or metal parts
Product Description
The scope of delivery for the high-pressure reactors BR-1000, BR-1500 and BR-2000 includes rupture disc, dip tube, valve and a tool for a proper opening and closing of the reactors. According to customer requests, the reactors can be manufactured with stainless steel 316 Ti (1.4571) or hatelloy C-22 (2.4602). Pressure measurement is made analog and/or digital.
Due to an extensive range of accessories, individual reactor configurations can be made easily. Examples of accessories are:
- Different heating methods (integrated or via additional instrument)
- Various stirring concepts
- Sample holders
- A number of control units
The BR-1000, BR-1500 and BR-2000 reactors can be used for volumes up to 2,230 ml and convince by an easy handling, which represents a high level of user friendliness. Amongst others, this can be traced back to the opening and closing tool. All reactor vessels can be equipped with PTFE inserts.
The reactors are used for different types of synthesis, material tests and catalysts researches. Depending on the users’ needs, two of the six ports (8 mm tube connection) can be used individually. When adding t-connections, the number of free ports can be enlarged.
Accessories
Constant heating or cooling by means of fluid.
Monitoring and controlling of heating for all Berghof reactors.
Electric heating jacket for consistent heating processes.
For Berghof reactors BR-300 to BR-200 as well as NR-1500 and NR-3000
For Berghof reactors BR-300 to BR-2000 as well as NR-1500 and NR-3000.
Suitable for all Berghof reactors.
Suitable for all Berghof reactors.
Berghof offers different types of catalysis baskets for applications in the field of catalysis research
Application Notes
Test samples of elastomers are prepared in coolants for tensile tests using Berghof reactors.
Raw and starting materials can be tested by the use of Berghof BR high-pressure reactors.
Berghof reactors for hydrothermal syntheses or for both sol-gel synthesis and hydrothermal synthesis.