Reactor Technology

NR-1500/3000

Low-pressure reactor

The low-pressure reactors NR-1500 and NR-3000 convince by a wide range of individual configuration options. They have volumes up to 3,650 ml and can be used for temperatures up to 300 °C and for pressures up to 25 bar. Furthermore, the reactors can be equipped with PTFE inserts. Typical applications of the NR-1500 and NR-3000 low pressure reactors are, amongst others, synthesis and material tests.

 Send inquiry
Temperature max.300 °C
Pressure max.25 bar
Volume max.3650 ml
Berghof low-pressure reactors NR-1500/NR-3000
more details

Technical Specifications: NR-1500/3000

NR-1500

NR-1500

NR-3000

NR-3000

General informationMaterialStainless Steel 316Ti (1.4571) or Hastelloy C-22 (2.4602)
Temperature max.300° C300° C300° C300° C
Pressure max.25 bar25 bar25 bar25 bar
Reactor VesselVolumeapprox. 1,900 mlapprox. 1,900 mlapprox. 3,650 mlapprox. 3,650 ml
Inner Diameter123 mm123 mm123 mm123 mm
Inner Height160 mm160 mm308 mm308 mm
Weightapprox. 5.7 kgapprox. 5.7 kgapprox. 8.1 kgapprox. 8.1 kg
Bottom Drain Valve--
TFM™-PTFE Insert (optional)*Volume-approx. 1,400 ml-approx. 2,900 ml
Inner Diameter-114 mm-114 mm
Inner Height-142 mm-290
ArmaturesStandard ArmaturesRupture Disc, Dip Tube, Valve, Tool
Pressure Measurementanalog and/or digital
Ports (total)7777
Ports (free)3**3**3**3**
Type of connection8 mm Tube Connection
Heating Systems (optional)Electrical (Heating Jacket)
Via Fluid (Double Jacket)
Via Fluid (Heating Coil)
Stirring (Magnetic Drive)RV-100
RV-400----

* When using PTFE inserts the maximum allowable temperature is 230° C
** The number of free ports can be enlarged by the use of T-connections

With our Product Comparison you can easily compare this product with others.
Compare products!

Range of Applications

Technology

  • Syntheses like hydrogenation and production of emulsions

Automotive

  • Material testing like aging tests at low pressures

Product Description

The NR-1500 and NR-3000 low-pressure reactors are characterized by an easy handling and, therefore, by a high level user-friendliness. E.g., for each reactor configuration, a stand is available. The reactors can be manufactured both of stainless steel and hastelloy.

According to customer needs, the reactors can be equipped with various accessories and extensions. These include the following three different heating systems:

  • Electrical heating via heating jacket
  • Double jacket with fluid
  • Heating coil with fluid

In addition, a cooling coil, stirring and PTFE inserts can be added to the reactors. All these components are retrofittable. A bottom drain valve can be fitted on the reactor, as well. However, reactors with a bottom drain valve cannot be equipped with PTFE inserts.

The NR-1500 and NR-3000 low-pressure reactors are often used in the automotive industry or in the technology sector. The applications are, amongst others, hydrogenations, material tests and production of emulsions at a low-pressure level.

Accessories

Accessory

BTC-3000 Temperature ControllerFor controlled heating processes

Monitoring and controlling of heating for all Berghof reactors.

 See accessory
Accessory

BTC-1000 Temperature ControllerFor controlled heating processes

Effective heating processes for both Berghof reactors and pressure vessels.

 See accessory
Accessory

BRM-1 Stirrer DriveFor a consistant stirring at up to 2,000 rpm

For Berghof reactors BR-300 to BR-200 as well as NR-1500 and NR-3000

 See accessory
Accessory

BRM-2 Stirrer DriveFor consistant stirring at up to 500 rpm

For Berghof reactors BR-300 to BR-2000 as well as NR-1500 and NR-3000.

 See accessory
Accessory

BDL-3000 Data LoggerFor reliable recording of temperature and pressure profiles

Suitable for all Berghof reactors.

 See accessory
Accessory

Sample HoldersFor material and corrosion testing

Suitable for all Berghof reactors.

 See accessory

Application Notes

Material Testing of Elastomers in Coolants

Test samples of elastomers are prepared in coolants for tensile tests using Berghof reactors.

 See application report

Material Testing of Elastomers in Blow-by-mixtures

Using Berghof reactor technology, elastomers can be prepared in blow-by-mixtures for tensile tests.

 See application report

Go back

MENU